FDA remembers iOS app software program that injured over 200 insulin pump customers Leave a comment


Not less than 224 folks with diabetes have reported accidents linked to a faulty iOS app that precipitated their insulin pumps to close down prematurely, in line with the US Meals and Drug Administration (FDA).

On Wednesday, the company introduced that California-based medical gadget producer Tandem Diabetes Care has issued a recall for model 2.7 of the iOS t:join cellular app, which is used along with the corporate’s t:slim X2 insulin pump. Particularly, the recall pertains to a software program situation that may trigger the app to repeatedly crash and relaunch, ensuing within the pump’s battery being drained by extreme Bluetooth communication.

This battery drain may cause the pump to close down “sooner than sometimes anticipated” in line with Tandem, although the pump will notify customers of an imminent shutdown through an alarm and low-power alert. The corporate has notified prospects to replace the cellular app to model 2.7.1 or later, which ought to repair the faulty software program.

The FDA has recognized this as a Class I recall — probably the most severe kind

Whereas no bodily recall is going down, the FDA has recognized this as a “Class I” recall — probably the most severe kind, because it pertains to points with merchandise that may probably trigger severe accidents or demise.

“Pump shutdown will trigger insulin supply to droop, which might result in an under-delivery of insulin and should lead to hyperglycemia, together with extreme hyperglycemia,” Tandem warned in its letter to prospects in March. “In extreme instances of hyperglycemia resulting from a protracted interval of no insulin supply, the consumer might expertise diabetic ketoacidosis and should require hospitalization or intervention from a medical skilled.”

No deaths linked to the problems have been reported as of April fifteenth. Tandem is encouraging pump customers to take explicit care after they sleep because it’s simpler to overlook battery depletion warnings, and is asking impacted prospects to substantiate they’ve been notified of the recall through this on-line type. For some other questions or considerations in regards to the insulin pump recall, prospects ought to contact Tandem Diabetes Care immediately.